Analytical Method Validation
When to use
Any time the question is whether an analytical procedure is fit for its intended purpose: designing a validation study, evaluating validation data, verifying a compendial procedure, transferring a procedure to another laboratory or instrument, or defending any of these in a report.
The two rules
1. Establish which framework governs before designing anything. The same assay validates differently under ICH Q2(R2), USP <1225>, ICH M10, CLSI EP, and ISO/IEC 17025. They differ in which characteristics are required, how the studies are laid out, and whether numeric acceptance criteria are supplied at all. Blending them produces a protocol that satisfies none of them.
2. State acceptance criteria before collecting data. Criteria chosen after seeing results are not acceptance criteria, and deciding them post hoc is a standing audit finding. ICH Q2(R2) deliberately supplies almost no numeric criteria — they have to come from the specification, the analytical target profile (ICH Q14 section 3), or development data. ICH M10 is the exception: it supplies explicit numbers, and they differ between chromatographic assays and ligand bin…